Modular laminar airflow system fbb 120 portal
Application
Laminar airflow modules can be used everywhere it is necessary to do working operations in dust-free surroundings (for example in health services, pharmaceuticals, precision mechanics, optics, electronics, etc.).
Individual modules can be assembled in rows creating continual working spaces with laminar airflow above work desks or technological lines. The work line is thus protected from particle and bacterial contamination.
The modules are intended for applications requiring laminar air flow for the protection of a product from particle and bacterial contamination. Substances which can endanger a worker’s health may not get released during product handling and processing.
- Fully automatic operation controlled by a microprocessor unit providing optimal parameters of the laminar airflow in working areas
- Low acquisition costs when compared with the typical clean room
- Using maintenance-free fans, which provide maximum reliability even during non-stop operation
- Assembly form enabling transport of the device through standard access routes
- Upon request the modular system can be made mobile
- In case of replacing the technology, the elements can be re-used for covering another technology
Construction
The module is equipped with a fan featuring an electronically-controlled electric engine. The fan revolutions are controlled by a microprocessor unit, which monitors the laminar flow velocity for each module separately and maintains the module in the programmed value. The monitored velocity is at the same time shown on the control unit display located on the portal control panel.
The standard velocity of the portals delivered is programmed to 0.45 m/s. Upon request, the velocity can be adjusted within the range of 0.2 to 0.6 m/s. A service engineer can make this programmed adjustment upon request also to a portal already delivered or in operation.
Modules are equipped with high-capacity fans with continuous automatic regulation of revolutions. This enables long-term maintenance of the required airflow velocity with high preciseness, despite gradual sedimentation of the filter, and thus optimal use of the module's maximum service life is possible.
The module sheath is made from zinc-coated metal sheet, whose exterior side is coated with polyester enamel, shade RAL 9010. A high-quality silencer guarantees silent operation. The air is filtered with a filtration element - in standard version with the filtration class H13 (HEPA), upon request a higher filtration class can be implemented. The frame is made from steel closed sections equipped with powder stoving enamel of a high chemical resistance.
Cleanlines level
Cleanless level according to EN ISO 14 644-1: ISO Class 5, according to US FS 209E: 100, in farmaceutical industry (Czech republic) according to VYR-7 (SUKL): class A
Accessories
Plastic lamellar curtains
The assembly of laminar modules may be delimited along its perimeter with overlapping plastic lamellar curtains (lamella width 200mm, thickness 2mm).
Lighting
Light in the working area is provided both from the outside through plastic curtains and by fluorescent lamps placed under the front panel of the portal.
Mobile version
Upon request, the module system PORTAL 2, 3, and 4 can be made as mobile.
The integrated version of the ceiling
Portal version
Standard dimension lines
| Type |
Woring area /mm/ |
Fan input at the velocity of 0.45 m.s-1 |
|
| width | depth | ||
| PORTAL 1 | 1160 | 565 | 380 |
| PORTAL 2 | 1160 | 1130 | 760 |
| PORTAL 2S | 2320 | 565 | 760 |
| PORTAL 3 | 1695 | 1160 | 1140 |
| PORTAL 4 | 2260 | 1160 | 1520 |
| PORTAL 8 | 2320 | 2355 | 3040 |
| PORTAL 10 | 2900 | 2355 | 3800 |